Customer Service: 010 541 0652

MX1102 - HOBO Bluetooth Carbon Dioxide, Temperature & Relative Humidity Data Logger

R10,510.08

Onset’s HOBO MX1102 CO2 logger makes it more convenient than ever to measure and record CO2 in buildings and other non-condensing environments. It measures CO2 from 0 – 5,000 parts per million (ppm) –..

  • Brands ONSET
  • Product Code: MX1102
  • Availability: 7-10 Days

Tags: CO2, temperature, humidity, data logger, MX1102, Bluetooth

MX1102 - HOBO Bluetooth Carbon Dioxide, Temperature & Relative Humidity Data Logger

Onset’s HOBO MX1102 CO2 logger makes it more convenient than ever to measure and record CO2 in buildings and other non-condensing environments. It measures CO2 from 0 – 5,000 parts per million (ppm) – and our free HOBOmobile app* allows you to access data right from your mobile phone or tablet within a 100-foot range.

The MX1102 also features a USB port so it can be used with a computer running HOBOware graphing and analysis software. 

 

Highlighted Features

  • Wireless communication via Bluetooth Low Energy (BLE) Technology
  • Six-month battery life at a 5-minute CO2 logging rate
  • Easy to deploy and offload using free HOBOmobile App
  • Visual and audible high & low alarm thresholds
  • Self-calibrating NDIR CO2 sensor technology ensures optimal accuracy and lower maintenance costs
  • Patented connectivity technology

 

In what environment does this data logger operate?

This data logger operates in an indoor environment.

 

What measurements does this data logger support?

 

The MX1102A data logger supports the following measurements: Carbon Dioxide, Relative Humidity and Temperature

 

For more information visit Onset

  •  
    HOBO MX1102 CO2 Data Logger
  • Four AA 1.5 V alkaline batteries
CO2 Sensor
CO2 Sensor Accuracy ±50 ppm ±5% of reading at 25°C (77°F), less than 90% RH non-condensing and 1,013 mbar
CO2 Sensor Calibration Auto or manual to 400 ppm
CO2 Sensor Pressure Dependence 0.13% of reading per mm Hg (corrected via user input for elevation/altitude)
CO2 Sensor Range 0 to 5,000 ppm
CO2 Sensor Warm-up Time 15 seconds
Compensated Pressure Range -305 to 5,486 m (-1,000 to 18,000 ft)
Non-linearity <1% of FS
Operating Pressure Range 950 to 1,050 mbar (use Altitude Compensation for outside of this range)
Sensing Method Non-dispersive infrared (NDIR) absorption
Logger
Download Type USB 2.0 interface or via Bluetooth Smart
Full Memory Download Time 20 seconds via USB; approximately 60 seconds via Bluetooth Smart, may take longer the further the device is from the logger
LCD LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range
Logger Battery Life 6 months, typical with logging and sampling intervals of 5 minutes or slower; 6 months or less with logging and sampling intervals faster than 5 minutes while logging CO2. Entering burst logging mode will impact battery life. With HOBOmobile use, battery life can be reduced by remaining connected, excessive readouts, checking of Full Status Details, audible alarms, and paging. Visual/audible alarms and other events can have a marginal impact on battery life.
Logger Battery Type 4 AA 1.5 Volt batteries (user replaceable) or USB power source (5 V DC, 2 Watts)
Logger CE The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).
Logger Environmental Rating IP50
Logger Memory 128 KB (84,650 measurements, maximum)
Logger Radio Power 1 mW (0 dBm)
Logger Time Accuracy ± 1 minute per month at 25°C (77°F)
Logger Transmission Range Approximately 30.5 m (100 ft) line-of-sight
Logger Weight 267.4 g (9.43 oz)
Logger Wireless Data Standard Bluetooth Smart (Bluetooth Low Energy, Bluetooth 4.0)
Logging Modes Fixed interval (normal, statistics) or burst
Logging Rate 1 second to 18 hours
Memory Modes Wrap when full or stop when full
Start Modes Immediate, push button, date & time, or next interval
Stop Modes When memory full, push button, date & time, or after a set logging period
Relative Humidity Sensor
Relative Humidity Sensor Accuracy ±2% from 20% to 80% typical to a maximum of ±4.5% including hysteresis at 25°C (77°F); below 20% and above 80% ±6% typical
Relative Humidity Sensor Drift <1% per year typical
Relative Humidity Sensor Range 1% to 90% RH (non-condensing)
Relative Humidity Sensor Resolution 0.01%
Response Time
Response Time Temperature 12 minutes to 90% in airflow of 1 m/s (2.2 mph)
Temperature Sensor
Temperature Sensor Accuracy ±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F)
Temperature Sensor Drift <0.1°C (0.18°F) per year
Temperature Sensor Range 0° to 50°C (32° to 122°F)
Temperature Sensor Resolution 0.024°C at 25°C (0.04°F at 77°F)